Features:
- Rugged: Dash 3000 is specially built to withstand the punishment of your most demanding intra-hospital transport applications with certifications for both ground and air transport
- Compact: Weighing no more than 12 pounds and measuring only 8.5″ x 10.5″ x 8.0″ regardless of configuration, these compact monitors make transport quick and trouble free
- Battery Technology: Dash 3000 has two user exchangeable batteries which provide 4-5 hours of run time. Battery status of each battery is displayed on screen continuously.
- Local Algorithms: All monitoring algorithms run locally on the Dash 3000 including full arrhythmia, multi-lead ST and interpretive 12 lead ECG. No need to purchase a separate piece of equipment.
- Comprehensive Parameter Set: Dash offers all the parameters needed to ensure that transporting patients is efficient as well as safe. Only Dash offers such a comprehensive set of parameters in a transport monitors.
- Motion Tolerant Algorithms: Multi-lead ECG analysis allows Dash to monitor all leads simultaneously reducing the affect of noise. DINAMAP NIBP technology ensures the most accurate reading at all times.
- Networking: The Unity network connection enables Dash to be admitted from the central station, view another patient, send a 12 lead report directly to the hospital MUSE cardiology management system and much more.
- Wireless LAN: Due to the superior Unity network, all the productivity saving network features available on Dash are maintained. There is no loss of capability when switching from hardwire to wireless.
- DASH PORT Docking Station: Quickly secure any Dash monitor with one simple operation seamlessly providing a power and a network connection.
Specifications:
Physical Weight & Dimensions
- Height: 36 cm
- Width: 28 cm
- Depth: 20 cm
- Weight: 5.5 kg
Display
- Size: 22 cm (diagonal).
- Type: active-matrix color TFT
- Resolution: 640 x 480 pixels
- Number of traces: 6 maximum
- Number of seconds/trace: 4.9 at 25 mm/sec
- Sweep Speed: 6.25, 12.5, 25 mm/sec (with erase bar)
- Information window: displays non-real-time information without obstructing the display of real-time information. Display organization: prioritized by parameter.
Controls
- Five hard keys: Silence Alarm, Graph G/Stop, NIBP Go/Stop, function (zero all) and Power On/Off.
- Trim Knob® control
Alarms
- Categories: patient status and system status
- Priority: 4-levels crisis, warning, advisory, message
- Notification: audible and visual. Settings: default or user-defined
- Silencing: 1 minute, current alarm only. Pause: 5 minutes in adult ICU mode, 3 minutes in Neonatal ICU mode, and 5, 15, minutes or permanent pause in OR mode.
ECG
- Standard leads available: I, II, III, V, aVR, aVL and aVF
- Leads analyzed simultaneously: I, II, III and V (multi-lead)
- Lead fail: identifies failed lead
- Alarms: user-selectable upper and lower heart rate limits
Respiration
- Range respiration rate: 0-200 breaths per minute. Use-selectable upper and lower respiration rate limits, and user-selectable apnea limit
- Measurement technique: impedance variation detection
Temperature
- Number of channels: 2
- Resolution: ±0.1°C
- Alarms: user-selectable upper and lower limits for T1, T2
IBP
- Number of channels: 2
- Range: -25mmHg to 300mmHg
- Alarms: user-selectable upper and lower limits for systolic, diastolic, and mean pressures
NIBP
- Measurement technique: oscillometric
- Displayed parameters: systolic, diastolic, and mean pressures, time of last measurement
- Measurement modes: manual, auto, and stat in Adult ICU and OR modes, manual and auto in neonatal mode
- Systolic range: 30-275 mmHg adult, 30-235 mmHg pediatric, 30-135 mmHg neonate
- Diastolic range: 10-220mmHg adult, 10-220mmHg pediatric, 10-110 mmHg neonate
Pulse Oximeter
- Parameters monitored: SpO2 and peripheral pulse rate
- SpO2 range: 50-100%, PPR range: 30-300 beats per minute
- SpO2 accuracy: ±2%, 70-100% SpO2,±3% (50-69% SpO2)
- PPR Accuracy: ±3 beats per minute