Description
The CASMED® 740 monitor offers leading-edge technology in a rugged, portable, lightweight unit that weighs only 1.4 kg (3 lbs).
Easy to use, reliable, accurate, with multi-parameter capabilities
- Fast, accurate blood pressure readings for neonates through adults . . . reduces the problems of motion in blood pressure measurement
- Monitor oximetry and temperature either continuously or in a “spot-check” mode
- Ideal for medical/surgical patients on pain medications including PCA pumps, endoscopy procedures, and for the EMS market
Featured Technologies
- CASMED’s proprietary MAXNIBP® technology
Oximetry choices include
- Masimo SET®
- Nellcor® OxiMax®
- Nonin®
Four models available
- CAS 740-1: MAXNIBP
- CAS 740-2: MAXNIBP and SpO2
- CAS 740-2T: MAXNIBP and Temperature
- CAS 740-3: MAXNIBP, SpO2 and Temperature
Features:
- Created with emergency medical responders in mind
- Wireless (infrared) printer option
- User removable NiMH battery pack, internal power supply and DC power option
- Flexible mounting solutions include cart, IV pole, bedrail, or wall; it can be fixed to a swivel wall mount designed specifically for emergency vechicles or carried in a protective, durable case
Specifications:
Physical Weight & Dimensions
- Height: 6.75 inches (17 cm)
- Width: 8.5 inches (21.5 cm)
- Depth: 3 inches (7.5 cm)
- Weight: 3 kg (1.4 lbs))
NIBP
- Technique: Oscillometric (MAXNIBP® Technology)
- Patient Range: Adult – Neonate
- Measurement Range:
- Systolic: 30-255 mmHg
- Diastolic: 15-220 mmHg
- MAP: 20-235 mmHg
- Pulse Rate Range: 30-240 BPM
- Accuracy:
- Blood Pressure: ± 5 mmHg standard deviation no greater than 8 mmHg
- Pulse Rate: ± 2% or ±2 BPM, whichever is greater
Oximetry (Options)
- Masimo SET®
- SpO2 Range: 0-100%
- SpO2 Accuracy:
- 70-100%, ± 2 digits (1 S.D.) Adult
- 70-100%, ± 3 digits (1 S.D.) Neonate
- Pulse Rate Range: 25-240 BPM
- Pulse Rate Accuracy: ± 3 BPM
- Nonin®
- SpO2 Range: 0-100%
- SpO2 Accuracy:
- 70-100%, ± 2 digits (1 S.D.) Adult
- 70-100%, ± 3 digits (1 S.D.) Neonate
- Pulse Rate Range: 18-240 BPM
- Pulse Rate Accuracy: ± 3% or ±1 digit, whichever is greater
- Nellcor® OxiMax®
- SpO2 Range: 1-100%
- SpO2 Accuracy:
- 70-100%, ± 2 digits (1 S.D.) Adult
- 70-100%, ± 3 digits (1 S.D.) Neonate
- Pulse Rate Range: 20-240 BPM
- Pulse Rate Accuracy: ± 3 digits
Temperature (Optional)
- Welch Allyn SureTemp® Plus:
- Temperature Range: 29 °C to 42 °C (84.2 °F to 107.6 °F)
- Accuracy: ± 0.1 °C (± 0.2 °F)
- Meets or exceeds ASTM Standards
Control Panel
- Display: LED display of measurement results, instructions, troubleshooting messages, and signal strength bar
- Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate, %SpO2, Temperature (°C or °F)
Patient Alarms
Adjustable high and low alarms for Pulse Rate, SpO2, Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure
Operating Modes
- History: Review of previous measurements
- NIBP: Manual, STAT or Automatic (at preset intervals)
- %SpO2: Continuous Monitoring
- Temperature: Predictive or Continuous
Power
- Source: External line or internal battery
- AC Power: 100-240 VAC, 50-60 Hz, 0.5A; T1.25A L250V fuse rating
- DC Power (EMS Option): 12-30 VDC
- Battery: NiMH battery pack (removable)
- Charge Time: 4 hours
- Operating Time: Minimum 4 hours of continuous operation
- Leakage Current: 100 _A (maximum)
Safety Limits
- 145 mmHg (Neonate)
- If measurement time exceeds 120 seconds (Adults), 90 seconds (Neonate)
- If safety timer detects microprocessor failure
Operating Environment
- Operating Temperature: 0°C to 50°C (32°F to 122°F)
- Storage Temperature: -20°C to 65°C (-4°F to 149°F)
- Humidity: 15 – 95%, non-condensing
- Altitude: Up to 3048 meters (10,000 feet) if stored or used outside temperature and humidity ranges.
Standards
- ANSI/AAMI SP-10
- EN 60601-1, EN 60601-1-2, EN 865, EN 60601-2-30, EN 60601-2-49
- ETL Listed – UL 60601, CAN/CSA C22.2 No. 601.1
- CE Marking according to Directive 93/42/EEC