Description
The top-of-the-line ERBOTOM ICC 350 heads ERBE’s family of units for HF surgery. They’ve given it one-of-a-kind functionality for all-round use, with the features and abilities developed in close teamwork with leading surgeons from many different medical disciplines.
The ICC 350 has received world-wide attention and recognition; it is a step forward from conventional surgical methods and opens the way for new ones.
Features:
Cutting
- Cut control – for automatic perfect cuts
- High cut with specialized arc regulation for underwater cuts or in fatty structures
- Endo cut – the cutting mode for endoscopy
- Mono- und biopolar cutting (Bipolar Cut)
- Power Peak System (PPS) for intelligent support of initial incision and cutting
Coagulation
- Precisely controlled coagulation and no carbonization with Soft Coag
- Forced Coag – for situations where speed is essential
- Spray Coag – for efficient surface coagulation
- Argon-Plasma Coagulation unit ERBE APC 300 hookup as standard equipment
- Mono- und bipolar coagulation (Bipolar Coag)
- Twin Coag – for comfortable coagulation on two channels: connector and individual settings for two Coag instruments
- Auto Start and Auto Stop – for coagulation support
Handling
- Maximum user-friendly operation via 11 individually configurable programs
Safety
- Nessy the one-of-a-kind Neutral Electrode Safety System
- Double-safe cutting with Cut Control and Biopolar Cut
Specifications:
Cutting
- HF power adjustment: With up/down increment keys
- Form of HF voltage: Sine-shaped
- Nominal frequency of HF voltage: 330 kHz
- Cutting quality: 4 coagulation effects, selectable by key
- Constancy of the coagulation effects: Automatically regulated
- Support for initial incision and cutting: PPS, PowerPeakSystem
- Monopolar AUTO CUT: With automatic voltage control
- Monopolar HIGH CUT: with automatic arc control
- Nominal HF output: 300 watts at 500 ohms
- Peak HF power with PPS: 400 watts at 500 ohms
- Monopolar ENDO CUT: 200 watts at max 200 ohms
- BIPOLAR AUTO CUT: With automatic voltage control
- Nominal HF output: 100 watts at 500 ohms
Coagulation
HF power adjustment: With up/down increment keys.
Monopolar SOFT COAG
- Form of HF current: Sine-shaped
- Nominal frequency of HF voltage: 300 kHz
- Nominal HF output: 120 watts at 125 ohms
- End of coagulation: with Pedal or AUTO STOP
Monopolar FORCED COAG
- Form of HF current: Pulse-modulated
- Nominal frequency of HF voltage: 1 MHz
- Nominal HF output: 120 atts at 350 ohms
Monopolar SPRAY COAG
- Form of HF current: Pulse-modulated
- Nominal frequency of HF voltage: 1 MHz
- Nominal HF output: 120 watts at 500 ohms
Biopolar COA
- Form of HF current: Sine-shaped
- Nominal frequency of HF voltage: 330 kHz
- Nominal HF output: 120 watts at 125 ohms
- Start of coagulation: By foot pedal or AUTO START
- End of coagulation: By foot pedal or AUTO STOP
Safety Precautions
- Safety class: I, corresponds to DIN IEC 601-1
- Type: CF, corresponds to DIN IEC 601-1
- Software check of LF leakage current, Icritical: 0.05mA (only with capacitive grounding)
- Software check of HF leakage current, Icritical: 150 mA/300MA
- NESSY: Neutral electrode Safety System
- Neutral electrode switching: Switch-selectable capacitively grounded or with floating power
- Self-Check: Includes all safety-related subassemblies
- Dimensions (W x H x D): 410 × 152 × 368 mm
- Weight: 10.0 kg
Power Connection
- Rated power voltage: 240V/230V/115V/110V/100V± 10%
- Rated power frequency: 50/60 Hz
- Power current: 4.0 A at 230 – 240 V/8.0 A at 100-115 V
- Power consumption in Standby mode: 25 watts
- Power consumption at max. HF power: 620 watts 920 VA
- Current consumption in Standby mode: 150 mA at 230 – 240V/300 mA at 100-115 V
- Potential equalization terminal: Yes
- Power fuses: 2 slow burn, 4 A at 230 – 240 V/8 A at 100-115 V