Description
Leading-edge patient monitor features patented Colin NIBP technology
The BP-S510 has a sophisticated design that provides simple, user-friendly operation. Its advanced functionality makes it perfect for a wide variety of uses, including post-anesthesia care, ambulatory surgery, endoscopy, emergency rooms and more!
- Measures – ECG (3/5 lead), heart rate, Colin NIBP, Nellcor SpO2 and respiration. Available with temperature, two-channel IBP, EtCO2, and a printer.
- Compact – Conveniently compact and feature-packed, the BP-S510 is great for areas where space is at a premium.
- Lightweight – Weighing only 12.1 lbs, the BP-S510 is easy to carry from ward to ward. This lightweight device makes it easy to move to stepdown, PACU, RR, etc.
- Battery – Operation for Portability: Available with rechargeable batteries, the BP-S510 can be used during transport within your facility.
- Wide Viewing Angle – The BP-S510’s large LCD screen features a wide viewing angle. Each parameter is color-coded in bright and vivid colors, making the information easy to read even from a distance. You can also select from multiple views to highlight just the information you need.
Specifications:
Non-Invasive Blood Pressure (NIBP) – Patented Colin Technology
- Measurement technology Oscillometric method
- Measurement method Linear deflation method
- NIBP Measurement range
- Adult/pediatric mode:
- SYS 60 to 250mmHg
- MAP 45 to 235mmHg
- DIA 40 to 200mmHg
- Pulse rate 40 to 200bpm
- Pressure display range 10 to 300mmHg
- Pressure display accuracy Less than ±3mmHg
- Pulse rate 40 to 240bpm
- NIBP accuracy AAMI SP-10:2002+A1:2003
- Pulse rate accuracy ±2% or ±2 beats
ECG
- Leads 3/5 Lead
- Lead I, II, III, aVR, aVL, aVF, V (Chest Lead)
- Heart Rate
- Measurement Range 20 to 250bpm
- Accuracy ±3bpm or ±5% (whichever is greater)
Respiration (im RR)
- Measurement method Trans-thoracic impedance
- Leads RA to LA
- Range 0, 3 to 120 breaths/min
- Accuracy ±3 breaths/min
SpO2 (Pulse oximeter)
- Measurement method 2 wave length pulse wave type
- Range 1 to 100% SpO2
- Accuracy Adult/pediatric (When using SpO2 Disposable Sensor MAX-A) 70 to 100% SpO2 ±2 digits, 1 to 69% SpO2 unspecified
- Pulse Rate Range 20 to 250bpm
- Pulse Rate Accuracy ±3bpm
- Display update Less than 30 sec
Temperature Measurement (Optional)
- Measurement method Thermistor
- Probe Type Thermistor probe YSI 400 series (2 channels) (Probes not included)
- Range 15 to 45°C (59 to 113°F)
- Display Accuracy ±0.1°C (25 to 45°C) (77 to 113°F) ±0.2°C (15 to less than 25°C) (59 to less than 77°F)
- Probe Accuracy ±0.1°C (±0.18°F)
IBP (Optional)
- Measurement Method: Pressure Transducer Method
- Pressure Measurement Range – -50 to +300mmHG
- Pulse Rate Measurement Range – 20 to 250bpm
- Display Accuracy – ±3mmHG
- Standard – IEC60601-2-34:2000
CO2 (Option)
- Measurement method Non-dispersive Infrared Spectroscopy
- Range 0-100mmHg : 0-13.3kpa : 0-10%CO2
- Accuracy ±2mmHg or ±4% (whichever is greater)
- Respiration (awRR) Range 0 to 120 breaths/min
- Respiration (awRR) Accuracy ±1 breath/min
Recorder (Option)
- Type Thermal
- Weight 0.33 lbs. (150 g)
- Resolution 8 dot/mm
- Number of channels 1 to 2 channels
- Paper Type Thermal
- Paper Width 2 inches (50 mm)
- Paper Speeds 25.0mm/sec and 50.0mm/sec (1in/sec and 2in/sec)
Power Requirement
- AC Power 100-240V~50/60Hz, 0.5-1.3A
- Battery Ni-MH, 12V/3.8Ah (1 included)
Environmental
- Operating Temperature 10 to 40°C (50 to 104°F)
- Storage Temperature -20 to 60°C (-4 to 140°F)
- Operating Humidity 15 to 90% RH, non-condensing
Reviews