Skip to content Skip to footer

Siemens Healthineers Epoc System, Refurbished

Brand: Siemens Healthineers

Compare
Category: Product ID: 4926

Description

The Siemens Healthineers epoc® system is the best point of care testing device (POCT device) on the market right now. It’s one of the reasons we became an authorized Siemens distributor. We put a lot of effort into finding the best instruments for our community so you don’t have to, and this POC diagnostic device is it. 

This point of care testing device (POCT device) improves patient outcome, quality assurance, and patient testing workflows while transforming patient care delivery across healthcare pathways. The epoc® Blood Analysis System gives you laboratory-quality results here and now, right at the patient’s bedside.

A small, wireless, handheld point of care testing device allows you to bring the clinical laboratory with you:

  • To the desktop of your medical practice
  • In your air ambulance
  • To your field clinic
  • To quickly test results of other health products in development

The right point of care system for your family medical practice.

The epoc® system is a handheld, wireless point of care testing device to enable comprehensive blood sample analysis testing at the patient’s bedside on a single room temperature test card, with results in less than a minute.

  • Streamline patient-testing workflow.
  • Accelerate clinical decision-making.
  • Simplify inventory management.
  • Deliver real-time results and reporting.

It allows you to insource a lot of the patient care testing you’ve been outsourcing, getting a test result in seconds rather than hours or days. 

 

Features:

Streamline patient-testing workflow.

By moving testing to the patient’s side, caregivers, patients, and hospital administration all benefit from a less-complex testing process that improves turnaround times, patient care, and operational efficiencies.

Accelerate clinical decision making.

Empower clinicians to make faster decisions with lab-quality results at the patient’s side, enhancing clinical decision making and improving patient outcomes.

Simplify inventory management.

By incorporating a full menu on a single, room temperature-stable test card bar-coded for quality assurance, the epoc system delivers a more efficient and easy-to-manage patient-side testing program.

Deliver real-time results and reporting.

With an integrated, wireless connection, you can easily transmit results from the patient’s side to the EMR/LIS/HIS in real time and gain centralized control of decentralized testing.

Specifications:

Measured Parameters

 

 

 

Parameter

Unit of Measure

 

Measurement Range

 

 

 

 

pH

pH units

 

6.5–8.0

pCO2

mmHg

 

5-250

 

kPa

 

0.7-33.3

pO2

mmHg

 

5–750

 

kPa

 

0.7–100

TCO2

 

mmol/L

5-50

 

 

mEq/L

 

Na+

mmol/L

 

85–180

 

mEq/L

 

 

K+

mmol/L

 

1.5–12.0

 

mEq/L

 

 

Ca++

mmol/L

 

0.25–4.00

 

mg/dL

 

1.0–16.0

 

mEq/L

 

0.5–8.0

Cl-

mmol/L

 

65–140

 

mEq/L

 

 

Hct

 

% PCV

10–75

 

 

L/L

0.10–0.75

Glu

mmol/L

 

1.1–38.5

 

mg/dL

 

20-700

 

g/L

 

0.20-7.00

Lac

mmol/L

 

0.30–20.00

 

mg/dL

 

2.7–180.2

 

g/L

 

0.03–1.80

Crea

mg/dL

 

0.30–15.00

 

μmol/L

 

27–1326

Hct

% PCV

 

10–75

 

L/L

 

0.10-0.75

BUN

mg/dL

 

3-120

 

 

 

 

Urea

mmol/L

 

1.1-42.8

 

mg/dL

 

7-257

 

g/L

 

0.07-2.57

Calculated Parameters

 

 

Parameter

Unit of Measure

Measurement Range

cHgb

mmol/L

2.0-15.5

 

g/dL

3.3-25

 

g/L

33-250

cHCO3

mmol/L

1-85

 

mEq/L

 

cTCO2

mmol/L

5-50

 

mEq/L

 

BE(ecf)

mmol/L

-30 – +30

 

mEq/L

 

BE(b)

mmol/L

-30 – +30

 

mEq/L

 

cSO2

%

0-100

GFRmdr*

mL/min/1.73m2

2–60 or >60

GFRmdr-a*

mL/min/1.73m2

2–60 or >60

GFRckd

mL/min/1.73m2

1-225

GFRckd-a

mL/min/1.73m2

1-225

GFRswz§

mL/min/1.73m2

1-275

AGap

mmol/L

-14 – +95

 

mEq/L

 

AGapK

mmol/L

-10 – +99

 

mEq/L

 

BUN/Crea

mg/mg

0.2-400.0

Urea/Crea

mmol/mmol

0.8-1615.4

 

mg/mg

0.4-856.8

A

mmHg

5-800

 

kPa

0.67-106.64

A-a

mmHg

1-800

 

kPa

0.13-106.64

a/A

%

0-100

 

fraction

0-1

Institutions should establish and set their own normal range values.
* Values >60 will be reported as >60 mL/min/1.73 m2.
† IDMS-traceable MDRD type
‡ CKD-EPI equation

Reviews